Instrumentation

We perform a lot of in-house analysis for our materials. This page showcases our own devices.

LUMOS - Learning-Driven Universal Material Optimization System

Flow Reactor

LUMOS is our platform, specifically designed for the continuous-flow synthesis and real-time optical characterisation of colloidal quantum dots (QDs), targeting bandgap energies ranging from approximately 0.5 to 3.10 eV. It combines modular flow processing, temperature-controlled reaction environments, and broadband in-line optical spectroscopy spanning the UV–visible to near-infrared regions. A custom-developed software interface enables comprehensive hardware integration, automated experimental control, real-time data acquisition, and autonomous optimisation of synthesis conditions using Bayesian optimisation. This allows target-driven exploration of reaction parameters and accelerates the development and optimisation of QD synthesis processes.

UV-Vis spectrometers

  • Avantes AvaSpec-ULS2048CL-EVO-RS
    • CMOS linear image sensor
    • Integration time: 9 µs - 59 s
    • Range: 200-1100 nm
    • Spectrometer slits available: 25, 50, 100 and 200 µm
  • Chemical Probes/ Fiber optics available:
    • 2 x QP400-2-SR-BX: 200-1100 nm range, 200 µm fiber optic core
    • QP600-1-XSR: 180-900 nm range, 200 µm fiber optic core
    • QR600-7-VIS/125F: 400-2100 nm, 600 µm fiber optic core
    • QP600-1-VIS-NIR: 400-2100 nm range, 600 µm fiber optic core
    • TP300-UV-Vis: 200-1100 nm range, 300 µm fiber optic core
  • 2 x Ocean Optics DH2000BAL: D2 and W lamp

Quantum HORIBA Scientific FluoroMax Plus Spectrofluorometer + Integrating Sphere Quanta-Phi

  • Excitation 330-nm blaze (220-600 nm optical range)
  • Emission 500-nm blaze (290-850 nm optical range)
  • Detector type: InGaAs, PMT
  • Lamps: Xe-lamp
  • f/3.5 Czerny-Turner design monochromator: 1200-grooves/mm gratings
  • Resolution 0.3 nm
  • Range: 0-950 nm (physical)
  • Step size: 0625-100 nm
  • PL range: 250-2500 nm – Detector type: InGaAs
  • 1 cm path cuvettes

UV-Vis-NIR spectrometer: Shimadzu UV-3600 i Plus

  • Range: 185-3300 nm, Resolution: 0.1 nm
  • Light source: 50W halogen lamp, D2 lamp
  • Grating-grating double monochromator
  • PMT detector: 185-1000 nm
  • InGaAs detector: 700-1800 nm
  • PbS detector: 1600-3300 nm
  • Film holder for thin films, glass slides or filters

Glovebox - MBRAUN UNIlab Plus

  • N2 flow circulation
  • Freezer: -15°C
  • Manual purge valve
  • Automatic Box pressure control with foot switch: -15 to +15 mbars

Single reactor system RKM: O2 removal by Cu-catalyst, H2O removal by adsorption on a proprietary adsorbent (molecular sieves) without heat exchanger

Glovebox - MBRAUN UNIlab Pro

  • N2 flow circulation
  • Glovebox Cooling function
  • Manual purge valve
  • Automatic box pressure control with foot switch: -15 to +15 mbars

Single reactor system RKM: O2 removal by Cu-catalyst, H2O removal by adsorption on a proprietary adsorbent (molecular sieves) without heat exchanger

Schlenk Lines

  • 6 lines using N2 flow
  • 3 lines using either Ar or N2 flow

3D printers - Ender 3 V3 SE + Saturn4 Ultra 16K

Shimadzu IRSpirit QATR-S

  • Range: 400-4000 cm-1
  • Diamond Crystal
  • Max resolution: 4 cm-1

Ossila Spin Coater

  • RPM range: 120-6000 rpm
  • 5x5 mm, 10 x 10 mm and 25x25 mm glass slides can be used. A different substrate can be 3D-printed to accommodate other glass slides dimensions

Centrifuges

  • Eppendorf miniSpin: 12 x 1.5/2 mL tubes, rpm: 800-13400, max rcf 14100 x g
  • Eppendorf 5425: 10 x 5 mL tubes, max rpm: 15060, max rcf 21300 x g
  • Eppendorf 5810: 6 x 50 mL tubes, max rpm: 11000, max rcf 20133 x g
  • Eppendorf 5418: 18 x 1.5/ 2 mL tubes, max rpm: 14000, max rcf 16873 x g

Sonica 2200 MH Sonication Bath

ThermoFisher Scientific Vacuum Oven VT6025

Mass Flow controller: Bronkhorst EL-FLOW-SELECT

  • Flow rate: 51325 - 100 scc/min
  • The mass flow controller is calibrated to ammonia gas. This set up allows for gas liquid reactions.

Ammonia gas pressure regulator Linde HiQ300 R300/1-3SS

  • Pressure gauge: 0-6 bar
  • While this regulator can be used for various gases, it is specifically designed to be resistant to ammonia corrosion.
  • One valve is for opening up the ammonia flow, while the second one is for purging the reducer for optimum lifespan of the equipment.

In-Situ Photoluminescence setup

  • Spectrometer: AvaSpec-ULS2048CI, Grating: 200-1100 nm
  • Optical fiber: Ocean Optics QR600-7-VIS125F. The diameter of the fiber is 600 μ Reflection probe with 6 collecting fibers and 1 illuminating one, with the wavelength range of 400-2100 nm.
  • Light source: AVALIGHT-HPLED-385, 385 nm
  • Comments: The current set up allows for real-time monitoring of the photoluminescence in the visible range with maximum acquisition speed of 9 μs - 59 s, while the maximum ranges between - 20 - 300°C (for temperatures above 80°C a metal jacket is required).

Gas-Liquid reactions

  • Flow meter: Vapourtec SF-10. Flow range: 0.5 - 99.0 scc/min
  • Pressure regulator: Linde HiQ R300/1-3SS. Pressure gauge: 0 - 6 bar
  • Comments: This set up can be combined with the in-situ PL set up to achieve in-situ photoluminescence of gas liquid reactions.